In endovascular interventions, thin, flexible instruments are inserted through the skin into the blood vessels to diagnose and treat various diseases of the vascular system. One drawback is that the instruments are difficult to maneuver in the desired direction due to limitations in shape and flexibility. Another disadvantage is that the interventions are performed under intermittent fluoroscopy/angiography imaging. Magnetic resonance imaging (MRI) may offer advantages over X-ray guidance. It presents a good soft tissue contrast without the use of nephrotoxic media or ionizing radiation. The aim of this study is to develop a guidewire that is compatible with MRI and includes a steerable segment at the tip. This added degree-of-freedom may improve the maneuverability of the devices thereby the efficiently and safety of the navigation. A 1.6 m (5 ft, 3 in.) long and 0.035 in. diameter guidewire that consists of MR compatible materials and has a flexible tip was designed. The only metallic part was a nitinol rod that was implemented at the distal flexible tip. To limit the risk of heating in the MRI, this rod was kept shorter than 30 mm. The tip could be deflected in one direction by pulling on a Dyneema wire that was placed in the lumen of the shaft of the guidewire. To drive the steerable tip, a handle that could be easily attached/detached from the instrument was designed and implemented. Using the handle, the tip of the 1.60 m long guidewire prototype could be actuated to reach angles from 30 deg to 250 deg. The handle could easily be placed on and removed from the guidewire, so conventional 0.035 in.–compatible catheters could slide over from the proximal end. However, in order to make the guidewire more efficient to enter a bifurcation, the stiffness of the tip should progressively increase from its proximal to its distal end. The guidewire was imaged in a 1.5T MRI using real-time imaging without producing artifacts that would have shaded the anatomy. It was possible to assemble a guidewire with a steerable segment in the required size, using MR compatible materials. Therefore, the current design is a promising proof of concept and allowed us to clearly identify the features that need to be improved in order to come to a clinically applicable instrument.
Skip Nav Destination
Delft University of Technology,
Delft 2628 CD,
e-mail: H.C.Clogenson@tudelft.nl
Delft University of Technology,
Delft 2628 CD,
Delft University of Technology,
Delft 2628 CD,
Article navigation
June 2014
Research-Article
Steerable Guidewire for Magnetic Resonance Guided Endovascular Interventions
H. C. M. Clogenson,
Delft University of Technology,
Delft 2628 CD,
e-mail: H.C.Clogenson@tudelft.nl
H. C. M. Clogenson
Department of Biomechanical Engineering
,Delft University of Technology,
Mekelweg 2
,Delft 2628 CD,
The Netherlands
e-mail: H.C.Clogenson@tudelft.nl
Search for other works by this author on:
J. Dankelman,
Delft University of Technology,
Delft 2628 CD,
J. Dankelman
Department of Biomechanical Engineering
,Delft University of Technology,
Mekelweg 2
,Delft 2628 CD,
The Netherlands
Search for other works by this author on:
J. J. van den Dobbelsteen
Delft University of Technology,
Delft 2628 CD,
J. J. van den Dobbelsteen
Department of Biomechanical Engineering
,Delft University of Technology,
Mekelweg 2
,Delft 2628 CD,
The Netherlands
Search for other works by this author on:
H. C. M. Clogenson
Department of Biomechanical Engineering
,Delft University of Technology,
Mekelweg 2
,Delft 2628 CD,
The Netherlands
e-mail: H.C.Clogenson@tudelft.nl
J. Dankelman
Department of Biomechanical Engineering
,Delft University of Technology,
Mekelweg 2
,Delft 2628 CD,
The Netherlands
J. J. van den Dobbelsteen
Department of Biomechanical Engineering
,Delft University of Technology,
Mekelweg 2
,Delft 2628 CD,
The Netherlands
Manuscript received June 12, 2013; final manuscript received January 17, 2014; published online March 7, 2014. Assoc. Editor: Carl A. Nelson.
J. Med. Devices. Jun 2014, 8(2): 021002 (7 pages)
Published Online: March 7, 2014
Article history
Received:
June 12, 2013
Revision Received:
January 17, 2014
Citation
Clogenson, H. C. M., Dankelman, J., and van den Dobbelsteen, J. J. (March 7, 2014). "Steerable Guidewire for Magnetic Resonance Guided Endovascular Interventions." ASME. J. Med. Devices. June 2014; 8(2): 021002. https://doi.org/10.1115/1.4026560
Download citation file:
Get Email Alerts
Related Articles
Time-Action Analysis of Catheter Manipulation During Navigation Tasks in Bifurcations
J. Med. Devices (December,2013)
A Modular Magnetic Resonance Imaging-Compatible Pipette Simulator to Evaluate How Design Affects the Basilar Thumb Joint Mechanics
J. Med. Devices (December,2022)
RFID-Based Real-Time Navigation for Interventional Magnetic Resonance Imaging: Development and Evaluation of a Novel Tracking System
J. Med. Devices (September,2017)
Smart Guidewires for Smooth Navigation in Neurovascular Intervention
J. Med. Devices (March,2015)
Related Proceedings Papers
Related Chapters
Using Statistical Learning Theory to Improve Treatment Response for Metastatic Colorectal Carcinoma
Intelligent Engineering Systems through Artificial Neural Networks, Volume 20
The Design of Location System Based on GPS
International Conference on Mechanical and Electrical Technology, 3rd, (ICMET-China 2011), Volumes 1–3
Mobile Robots for Earth Exploration: Applications, Technologies and Image Processing Techniques for Navigation
Mobile Robots for Dynamic Environments