Mechanical stimulation is essential for chondrocyte metabolism and cartilage matrix deposition. Traditional methods for evaluating developing tissue in vitro are destructive, time consuming, and expensive. Nondestructive evaluation of engineered tissue is promising for the development of replacement tissues. Here we present a novel instrumented bioreactor for dynamic mechanical stimulation and nondestructive evaluation of tissue mechanical properties and extracellular matrix (ECM) content. The bioreactor is instrumented with a video microscope and load cells in each well to measure tissue stiffness and an ultrasonic transducer for evaluating ECM content. Chondrocyte-laden hydrogel constructs were placed in the bioreactor and subjected to dynamic intermittent compression at 1 Hz and 10% strain for 1 h, twice per day for 7 days. Compressive modulus of the constructs, measured online in the bioreactor and offline on a mechanical testing machine, did not significantly change over time. Deposition of sulfated glycosaminoglycan (sGAG) increased significantly after 7 days, independent of loading. Furthermore, the relative reflection amplitude of the loaded constructs decreased significantly after 7 days, consistent with an increase in sGAG content. This preliminary work with our novel bioreactor demonstrates its capabilities for dynamic culture and nondestructive evaluation.
Skip Nav Destination
e-mail: jenni.popp@nist.gov
Article navigation
June 2012
Research Papers
An Instrumented Bioreactor for Mechanical Stimulation and Real-Time, Nondestructive Evaluation of Engineered Cartilage Tissue
Jenni R. Popp,
Jenni R. Popp
Materials Reliability Division,
e-mail: jenni.popp@nist.gov
National Institute of Standards and Technology
, 325 Broadway, Boulder, CO 80305
Search for other works by this author on:
Justine J. Roberts,
Justine J. Roberts
Department of Chemical and Biological Engineering,
University of Colorado
, Boulder, CO 80309
Search for other works by this author on:
Doug V. Gallagher,
Doug V. Gallagher
Materials Reliability Division,
National Institute of Standards and Technology
, 325 Broadway, Boulder, CO 80305
Search for other works by this author on:
Kristi S. Anseth,
Kristi S. Anseth
Department of Chemical and Biological Engineering,
University of Colorado
, Boulder, CO 80309
Search for other works by this author on:
Stephanie J. Bryant,
Stephanie J. Bryant
Department of Chemical and Biological Engineering,
University of Colorado
, Boulder, CO 80309
Search for other works by this author on:
Timothy P. Quinn
Timothy P. Quinn
Materials Reliability Division,
National Institute of Standards and Technology
, 325 Broadway, Boulder, CO 80305
Search for other works by this author on:
Jenni R. Popp
Materials Reliability Division,
National Institute of Standards and Technology
, 325 Broadway, Boulder, CO 80305e-mail: jenni.popp@nist.gov
Justine J. Roberts
Department of Chemical and Biological Engineering,
University of Colorado
, Boulder, CO 80309
Doug V. Gallagher
Materials Reliability Division,
National Institute of Standards and Technology
, 325 Broadway, Boulder, CO 80305
Kristi S. Anseth
Department of Chemical and Biological Engineering,
University of Colorado
, Boulder, CO 80309
Stephanie J. Bryant
Department of Chemical and Biological Engineering,
University of Colorado
, Boulder, CO 80309
Timothy P. Quinn
Materials Reliability Division,
National Institute of Standards and Technology
, 325 Broadway, Boulder, CO 80305J. Med. Devices. Jun 2012, 6(2): 021006 (7 pages)
Published Online: April 26, 2012
Article history
Received:
November 18, 2011
Revised:
February 29, 2012
Online:
April 26, 2012
Published:
April 26, 2012
Citation
Popp, J. R., Roberts, J. J., Gallagher, D. V., Anseth, K. S., Bryant, S. J., and Quinn, T. P. (April 26, 2012). "An Instrumented Bioreactor for Mechanical Stimulation and Real-Time, Nondestructive Evaluation of Engineered Cartilage Tissue." ASME. J. Med. Devices. June 2012; 6(2): 021006. https://doi.org/10.1115/1.4006546
Download citation file:
Get Email Alerts
Context-Driven Design of a Laparoscopic Instrument Cleaner for Use in Rural Low-Resource Hospitals
J. Med. Devices (March 2025)
Controlled Ice Nucleation With a Sand-PDMS Film Device Enhances Cryopreservation of Mouse Preantral Ovarian Follicles
J. Med. Devices (December 2024)
Review of Blood and Fluid Warming Methods
J. Med. Devices (December 2024)
Related Articles
A Finite Element Model of Cell-Matrix Interactions to Study the Differential Effect of Scaffold Composition on Chondrogenic Response to Mechanical Stimulation
J Biomech Eng (April,2011)
Design of an Endoreactor for the Cultivation of a Joint-Like-Structure
J. Med. Devices (June,2009)
Related Proceedings Papers
Related Chapters
Effects of Ultrasound Stimulation on Chondrocytes in Three-Dimensional Culture in Relation to the Production of Regenerative Cartilage Tissue
Biomedical Applications of Vibration and Acoustics in Therapy, Bioeffect and Modeling
Introduction and Scope
High Frequency Piezo-Composite Micromachined Ultrasound Transducer Array Technology for Biomedical Imaging
Tissue Harmonic Motion Estimation for Tissue Characterization Using Ultrasound and Kalman Filter
Biomedical Applications of Vibration and Acoustics in Imaging and Characterizations