A local heat–flux measurement system was built, calibrated, and tested for use in unsteady flows. The system was designed to maintain constant-wall-temperature boundary conditions. The measuring element is a thin-film heat flux gage made by sputter-coating gold onto a substrate. A constant-temperature anemometer is used to maintain the thin-film gage at a specified temperature under fluctuating conditions. A separate temperature control system maintains the surrounding boundary at the gage temperature. The system was calibrated for both steady and unsteady flows using a specially designed calibrator for local heat flux gages. The steady calibration was done with predominantly convective heat transfer. The unsteady calibration was achieved by adding oscillating radiant energy to the surface. Consequently, quantitative results can be obtained for both the mean and fluctuating components of the heat transfer. The frequency response was good to over 90 Hz. Sample results are presented of the unsteady heat transfer over a circular cylinder caused by natural vortex shedding at 70 to 80 Hz.
Skip Nav Destination
Article navigation
Research Papers
Design and Calibration of a Local Heat-Flux Measurement System for Unsteady Flows
D. S. Campbell,
D. S. Campbell
Mechanical Engineering Department, Virginia Polytechnic Institute and State University, Blacksburg, VA 24061
Search for other works by this author on:
M. Gundappa,
M. Gundappa
Mechanical Engineering Department, Virginia Polytechnic Institute and State University, Blacksburg, VA 24061
Search for other works by this author on:
T. E. Diller
T. E. Diller
Mechanical Engineering Department, Virginia Polytechnic Institute and State University, Blacksburg, VA 24061
Search for other works by this author on:
D. S. Campbell
Mechanical Engineering Department, Virginia Polytechnic Institute and State University, Blacksburg, VA 24061
M. Gundappa
Mechanical Engineering Department, Virginia Polytechnic Institute and State University, Blacksburg, VA 24061
T. E. Diller
Mechanical Engineering Department, Virginia Polytechnic Institute and State University, Blacksburg, VA 24061
J. Heat Transfer. May 1989, 111(2): 552-557 (6 pages)
Published Online: May 1, 1989
Article history
Received:
April 28, 1986
Online:
October 20, 2009
Citation
Campbell, D. S., Gundappa, M., and Diller, T. E. (May 1, 1989). "Design and Calibration of a Local Heat-Flux Measurement System for Unsteady Flows." ASME. J. Heat Transfer. May 1989; 111(2): 552–557. https://doi.org/10.1115/1.3250713
Download citation file:
Get Email Alerts
Cited By
Entropic Analysis of the Maximum Output Power of Thermoradiative Cells
J. Heat Mass Transfer
Molecular Dynamics Simulations in Nanoscale Heat Transfer: A Mini Review
J. Heat Mass Transfer
Related Articles
Oscillatory Heat Transfer in a Pipe Subjected to a Laminar Reciprocating Flow
J. Heat Transfer (August,1996)
Frequency Response Characteristics of an Active Heat Flux Gage
J. Heat Transfer (August,1998)
Bounds on Heat Transfer in a Periodic Graetz Problem
J. Heat Transfer (February,1991)
Laminar Forced Convection From a Circular Cylinder Placed in a Micropolar Fluid
J. Heat Transfer (March,2007)
Related Proceedings Papers
Related Chapters
Mixed-Up Convection
Hot Air Rises and Heat Sinks: Everything You Know about Cooling Electronics Is Wrong
Surface Analysis and Tools
Tribology of Mechanical Systems: A Guide to Present and Future Technologies
Transient Temperature and Heat Flux Measurement Using Thin-Film Microsensors
Ultrasonic Welding of Lithium-Ion Batteries