Strut thickness in coronary stents has been shown to affect their clinical performance. The purpose of this investigation was to understand how stent strut thickness influenced stress in the vessel wall. Finite element modeling was used to compare stress distribution between a thick and thin strut stent in a curved coronary artery model. Specifically, we modeled two 8 mm long Guidant MULTI-LINK VISION stents, one with strut thickness of 81 μm and the other with strut thickness of 112 μm. Results showed larger values of stress, both in the radial direction and in the circumferential direction, for the thicker stent model. As expected, these higher stress values occurred beneath the stent struts and at the edges of the stent. Stented vessel compliance measurements of both models revealed that the thick strut VISION stent is less compliant than the thin strut VISION stent. These results support the hypothesis that compliance mismatch between the stent and the host vessel contributes to excessive stress in the walls of stented vessels.
Skip Nav Destination
ASME 2004 International Mechanical Engineering Congress and Exposition
November 13–19, 2004
Anaheim, California, USA
Conference Sponsors:
- Bioengineering Division
ISBN:
0-7918-4703-9
PROCEEDINGS PAPER
Investigation of Strut Thickness in Coronary Stents via Finite Element Analysis
Saami K. Yazdani,
Saami K. Yazdani
Wake Forest University Medical Center
Search for other works by this author on:
Joel L. Berry
Joel L. Berry
Wake Forest University Medical Center
Search for other works by this author on:
Saami K. Yazdani
Wake Forest University Medical Center
Joel L. Berry
Wake Forest University Medical Center
Paper No:
IMECE2004-61873, pp. 51-52; 2 pages
Published Online:
March 24, 2008
Citation
Yazdani, SK, & Berry, JL. "Investigation of Strut Thickness in Coronary Stents via Finite Element Analysis." Proceedings of the ASME 2004 International Mechanical Engineering Congress and Exposition. Advances in Bioengineering. Anaheim, California, USA. November 13–19, 2004. pp. 51-52. ASME. https://doi.org/10.1115/IMECE2004-61873
Download citation file:
5
Views
0
Citations
Related Proceedings Papers
Estimate of Migration Force for Stent
IMECE2001
Related Articles
A Finite Element Study on Variations in Mass Transport in Stented Porcine Coronary Arteries Based on Location in the Coronary Arterial Tree
J Biomech Eng (June,2013)
Deployment of Interwoven Stents in an Artery With Moderate Stenosis
J. Med. Devices (June,2011)
Two-Dimensional Optimization of a Stent for an Aneurysm
J. Med. Devices (June,2010)
Related Chapters
mDFA Human Empirical Results
Modified Detrended Fluctuation Analysis (mDFA)
Subsection NB—Class 1 Components
Companion Guide to the ASME Boiler & Pressure Vessel Code, Volume 1, Second Edition
Subsection NB—Class 1 Components
Companion Guide to the ASME Boiler and Pressure Vessel Code, Volume 1, Third Edition