0
Technical Brief

Design of a Single Degree-of-Freedom, Adaptable Electromechanical Gait Trainer for People With Neurological Injury

[+] Author and Article Information
Sung Yul Shin

Department of Mechanical Engineering,
University of Texas at Austin,
204 E Dean Keeton St,
Austin, TX 78712
e-mail: syshin0228@utexas.edu

Ashish D. Deshpande

Mem. ASME
Department of Mechanical Engineering,
University of Texas at Austin,
204 E Dean Keeton St,
Austin, TX 78712
e-mail: ashish@austin.utexas.edu

James Sulzer

Department of Mechanical Engineering,
University of Texas at Austin,
204 E Dean Keeton St,
Austin, TX 78712
e-mail: james.sulzer@austin.utexas.edu

1Corresponding author.

Contributed by the Mechanisms and Robotics Committee of ASME for publication in the JOURNAL OF MECHANISMS AND ROBOTICS. Manuscript received August 17, 2017; final manuscript received March 21, 2018; published online May 31, 2018. Assoc. Editor: Veronica J. Santos.

J. Mechanisms Robotics 10(4), 044503 (May 31, 2018) (7 pages) Paper No: JMR-17-1258; doi: 10.1115/1.4039973 History: Received August 17, 2017; Revised March 21, 2018

The cost of therapy is one of the most significant barriers to recovery after neurological injury. Robotic gait trainers move the legs through repetitive, natural motions imitating gait. Recent meta-analyses conclude that such training improves walking function in neurologically impaired individuals. While robotic gait trainers promise to reduce the physical burden on therapists and allow greater patient throughput, they are prohibitively costly. Our novel approach is to design a new single degree-of-freedom (DoF) robotic trainer that maintains the key advantages of the expensive trainers but with a simplified design to reduce cost. Our primary design challenge is translating the motion of a single actuator to an array of natural gait trajectories. We address this with an eight-link Jansen mechanism that matches a generalized gait trajectory. We then optimize the mechanism to match different trajectories through link length adjustment based on nine different gait patterns obtained from gait database of 113 healthy individuals. To physically validate the range in gait patterns produced by the simulation, we tested kinematic accuracy on a motorized wooden proof-of-concept of the gait trainer. The simulation and experimental results suggested that an adjustment of two links can reasonably fit a wide range of gait patterns under typical within-subject variance. We conclude that this design could provide the basis for a low-cost, patient-based electromechanical gait trainer for neurorecovery.

FIGURES IN THIS ARTICLE
<>
Copyright © 2018 by ASME
Your Session has timed out. Please sign back in to continue.

References

Mozaffarian, D. , Benjamin, E. J. , Go, A. S. , Arnett, D. K. , Blaha, M. J. , Cushman, M. , Das, S. R. , de Ferranti, S. , Després, J. , Fullerton, H. J. , Howard, V. J. , Huffman, M. D. , Isasi, C. R. , Jiménez, M. C. , Judd, S. E. , Kissela, B. M. , Lichtman, J. H. , Lisabeth, L. D. , Liu, S. , Mackey, R. H. , Magid, D. J. , McGuire, D. K. , Mohler, E. R. , Moy, C. S. , Muntner, P. , Mussolino, M. E. , Nasir, K. , Neumar, R. W. , Nichol, G. , Palaniappan, L. , Pandey, D. K. , Reeves, M. J. , Rodriguez, C. J. , Rosamond, W. , Sorlie, P. D. , Stein, J. , Towfighi, A. , Turan, T. N. , Virani, S. S. , Woo, D. , Yeh, R. W. , and Turner, M. B. , 2016, “ Executive Summary: Heart Disease and Stroke Statistics-2016 Update: A Report From the American Heart Association,” Circulation, 133(4), pp. 447–454. [CrossRef] [PubMed]
Godwin, K. M. , Wasserman, J. , and Ostwald, S. K. , 2011, “ Cost Associated With Stroke: Outpatient Rehabilitative Services and Medication,” Top. Stroke Rehabil., 18(Suppl. 1), pp. 676–684. [CrossRef] [PubMed]
Colombo, G. , Joerg, M. , Schreier, R. , and Dietz, V. , 2000, “ Treadmill Training of Paraplegic Patients With a Robotic Orthosis,” J. Rehabil. Res. Dev., 37(6), pp. 693–700. https://www.rehab.research.va.gov/jour/00/37/6/abscolombo.htm
Banala, S. K. , Kim, S. H. , Agrawal, S. K. , and Scholz, J. P. , 2009, “ Robot Assisted Gait Training With Active Leg Exoskeleton (ALEX),” IEEE Trans. Neural Syst. Rehabil. Eng., 17(1), pp. 2–8. [CrossRef] [PubMed]
Veneman, J. F. , Ekkelenkamp, R. , Kruidhof, R. , Van der Helm, F. C. T. , and Van der Kooij, H. , 2005, “ Design of a Series Elastic-and Bowden Cable-Based Actuation System for Use as Torque-Actuator in Exoskeleton-Type Training,” Ninth International Conference on Rehabilitation Robotics (ICORR), Chicago, IL, June 28–July 1, pp. 496–499.
Hesse, S. , and Uhlenbrock, D. , 2000, “ A Mechanized Gait Trainer for Restoration of Gait,” J. Rehabil. Res. Dev., 37(6), pp. 701–708. https://www.rehab.research.va.gov/jour/00/37/6/abshesse.htm [PubMed]
Schmid, H. , Hesse, S. , Benhardt, R. , and Kruger, J. , 2005, “ Haptic Walker—A Novel HaPtic Foot Device,” ACM Trans. Appl. Percept., 2(2), pp. 166–180. [CrossRef]
Freivogel, S. , Mehrholz, J. , Husak-Sotomayor, T. , and Schmalohr, D. , 2008, “ Gait Training With the Newly Developed ‘LokoHelp’-System is Feasible for Non-Ambulatory Patients After Stroke, Spinal Cord and Brain Injury. A Feasibility Study,” Brain Inj., 22(7–8), pp. 625–632. [CrossRef] [PubMed]
Mehrholz, J. , and Pohl, M. , 2012, “ Electromechanical-Assisted Gait Training After Stroke: A Systematic Review Comparing End-Effector and Exoskeleton Devices,” J. Rehabil. Med., 44(3), pp. 193–199. [CrossRef] [PubMed]
Hesse, S. , Bertelt, C. , Jahnke, M. T. , Schaffrin, A. , Baake, P. , Malezic, M. , and Mauritz, K. H. , 1995, “ Treadmill Training With Partial Body Weight Support Compared With Physiotherapy in Nonambulatory Hemiparetic Patients,” Stroke, 26(6), pp. 976–981. [CrossRef] [PubMed]
Pohl, M. , Werner, C. , Holzgraefe, M. , Kroczek, G. , Mehrholz, J. , Wingendorf, I. , Hoölig, G. , Koch, R. , and Hesse, S. , 2007, “ Repetitive Locomotor Training and Physiotherapy Improve Walking and Basic Activities of Daily Living After Stroke: A Single-Blind, Randomized Multicentre Trial (DEutsche GAngtrainerStudie, DEGAS),” Clin. Rehabil., 21(1), pp. 17–27. [CrossRef] [PubMed]
Hesse, S. , Werner, C. , Uhlenbrock, D. , Frankenberg, S. V. , Bardeleben, A. , and Brandl-Hesse, B. , 2001, “ An Electromechanical Gait Trainer for Restoration of Gait in Hemiparetic Stroke Patients: Preliminary Results,” Neurorehabil. Neural Repair, 15(1), pp. 39–50. [CrossRef] [PubMed]
Werner, C. , Von Frankenberg, S. , Treig, T. , Konrad, M. , and Hesse, S. , 2002, “ Treadmill Training With Partial Body Weight Support and an Electromechanical Gait Trainer for Restoration of Gait in Subacute Stroke Patients a Randomized Crossover Study,” Stroke, 33(12), pp. 2895–2901. [CrossRef] [PubMed]
Mayr, A. , Kofler, M. , Quirbach, E. , Matzak, H. , Fröhlich, K. , and Saltuari, L. , 2007, “ Prospective, Blinded, Randomized Crossover Study of Gait Rehabilitation in Stroke Patients Using the Lokomat Gait Orthosis,” Neurorehabil. Neural Repair., 21(4), pp. 307–314. [CrossRef] [PubMed]
Simons, C. D. M. , van Asseldonk, E. H. F. , Folkersma, J. I. , van den Hoek, J. , Postma, M. , and Buurke, J. H. , 2009, “ First Clinical Results With the New Innovative Robotic Gait Trainer LOPES,” Gait Posture, 30(2), p. S7. [CrossRef]
Mehrholz, J. , Thomas, S. , Werner, C. , Kugler, J. , Pohl, M. , and Elsner, B. , 2017, “ Electromechanical-Assisted Training for Walking After Stroke,” Cochrane Database Syst. Rev., 5, p. CD006185.
Nam, K. Y. , Kim, H. J. , Kwon, B. S. , Park, J. W. , Lee, H. J. , and Yoo, A. , 2017, “ Robot-Assisted Gait Training (Lokomat) Improves Walking Function and Activity in People With Spinal Cord Injury: A Systematic Review,” J. Neuroeng. Rehabil., 14(1), p. 24. [CrossRef] [PubMed]
Husemann, B. , Müller, F. , Krewer, C. , Heller, S. , and Koenig, E. , 2007, “ Effects of Locomotion Training With Assistance of a Robot-Driven Gait Orthosis in Hemiparetic Patients After Stroke a Randomized Controlled Pilot Study,” Stroke, 38(2), pp. 349–354. [CrossRef] [PubMed]
Peurala, S. H. , Tarkka, I. M. , Pitkänen, K. , and Sivenius, J. , 2005, “ The Effectiveness of Body Weight-Supported Gait Training and Floor Walking in Patients With Chronic Stroke,” Arch. Phys. Med. Rehabil., 86(8), pp. 1557–1564. [CrossRef] [PubMed]
Westlake, K. P. , and Patten, C. , 2009, “ Pilot Study of Lokomat versus Manual-Assisted Treadmill Training for Locomotor Recovery Post-Stroke,” J. Neuroeng. Rehabil., 6(1), p. 18. [CrossRef] [PubMed]
Robotics Business Review, 2014, “ Healthcare Robotics: 2014, Leading Robotics Companies, Their Technology, Markets and Future,” Robotics Business Review, Framingham MA, accessed Feb. 21, 2015, https://www.roboticsbusinessreview.com/download/research-report-healthcare-robotics-2014/
Patrick, M. , 2014, “ Overview: Assessing Hospital Companies' Capital Expenditures, in an Investor's Guide to the US Hospital Industry PART 11 OF 16, Market Realist,” Market Realist Inc., New York, accessed, Nov. 16, 2014, https://marketrealist.com/2014/11/understanding-structure-us-hospital-industry
Freivogel, S. , Schmalohr, D. , and Mehrholz, J. , 2009, “ Improved Walking Ability and Reduced Therapeutic Stress With an Electromechanical Gait Device,” J. Rehabil. Med., 41(9), pp. 734–739. [CrossRef] [PubMed]
Beres-Jones, J. A. , and Harkema, S. J. , 2004, “ The Human Spinal Cord Interprets Velocity-Dependent Afferent Input During Stepping,” Brain, 127(10), pp. 2232–2246. [CrossRef] [PubMed]
Dietz, V. , Colombo, G. , and Jensen, L. , 1994, “ Locomotor Activity in Spinal Man,” Lancet, 344(8932), pp. 1260–1263. [CrossRef] [PubMed]
Laufer, Y. , Dickstein, R. , Chefez, Y. , and Marcovitz, E. , 2001, “ The Effect of Treadmill Training on the Ambulation of Stroke Survivors in the Early Stages of Rehabilitation,” J. Rehabil. Res. Dev., 38(1), pp. 69–78. https://www.rehab.research.va.gov/jour/01/38/1/laufer.html
Macko, R. F. , Ivey, F. M. , Forrester, L. W. , Hanley, D. , Sorkin, J. D. , Katzel, L. I. , Silver, K. H. , and Goldberg, A. P. , 2005, “ Treadmill Exercise Rehabilitation Improves Ambulatory Function and Cardiovascular Fitness in Patients With Chronic Stroke a Randomized, Controlled Trial,” Stroke, 36(10), pp. 2206–2211. [CrossRef] [PubMed]
Dietz, V. , 2002, “ Proprioception and Locomotor Disorders,” Nat. Rev. Neurosci., 3(10), pp. 781–790. [CrossRef] [PubMed]
Jansen, B. , Doubrovski, E. L. , and Verlinden, J. C. , 2014, “ Animaris Geneticus Parvus: Design of a Complex Multi-Body Walking Mechanism,” Rapid Prototyp. J., 20(4), pp. 311–319. [CrossRef]
Cunningham, D. A. , Rechnitzer, P. A. , Pearce, M. E. , and Donner, A. P. , 1982, “ Determinants of Self-Selected Walking Pace Across Ages 19 to 66,” J. Gerontol., 37(5), pp. 560–564. [CrossRef] [PubMed]
Koopman, B. , Van Asseldonk, E. H. F. , and Van der Kooij, H. , 2014, “ Speed-Dependent Reference Joint Trajectory Generation for Robotic Gait Support,” J. Biomech., 47(6), pp. 1447–1458. [CrossRef] [PubMed]
Yun, Y. , Kim, H. C. , Shin, S. Y. , Lee, J. , Deshpande, A. D. , and Kim, C. , 2014, “ Statistical Method for Prediction of Gait Kinematics With Gaussian Process Regression,” J. Biomech., 47(1), pp. 186–192. [CrossRef] [PubMed]
Norton, R. L. , 2004, Design of Machinery: An Introduction to the Synthesis and Analysis of Mechanisms and Machines, McGraw-Hill, New York.
Ravindran, A. , Reklaitis, G. V. , and Ragsdell, K. M. , 2006, Engineering Optimization: Methods and Applications, Wiley, Hoboken, NJ.
Kora, K. , Stinear, J. , and McDaid, A. , 2017, “ Design, Analysis, and Optimization of an Acute Stroke Gait Rehabilitation Device,” ASME J. Med. Devices, 11(1), p. 014503. [CrossRef]
Jung, C. Y. , Choi, J. , Park, S. , Lee, J. M. , Kim, C. , and Kim, S. J. , 2014, “ Design and Control of an Exoskeleton System for Gait Rehabilitation Capable of Natural Pelvic Movement,” IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS), Chicago, IL, Sept. 14–18, pp. 2095–2100.
Morone, G. , Paolucci, S. , Cherubini, A. , De Angelis, D. , Venturiero, V. , Coiro, P. , and Iosa, M. , 2017, “ Robot-Assisted Gait Training for Stroke Patients: Current State of the Art and Perspectives of Robotics,” Neuropsychiatr. Dis. Treat., 15(13), pp. 1303–1311. [CrossRef]
van Asseldonk, E. H. F. , Veneman, J. F. , Ekkelenkamp, R. , Buurke, J. H. , van der Helm, F. C. T. , and van der Kooij, H. , 2008, “ The Effects on Kinematics and Muscle Activity of Walking in a Robotic Gait Trainer During Zero-Force Control,” IEEE Trans. Neural Syst. Rehabil. Eng., 16(4), pp. 360–370. [CrossRef] [PubMed]
Duschau-Wicke, A. , von Zitzewitz, J. , Caprez, A. , Lunenburger, L. , and Riener, R. , 2010, “ Path Control: A Method for Patient-Cooperative Robot-Aided Gait Rehabilitation,” IEEE Trans. Neural Syst. Rehabil. Eng., 18(1), pp. 38–48. [CrossRef] [PubMed]
Riener, R. , 2016, “ Technology of the Robotic Gait Orthosis Lokomat,” Neurorehabilitation Technology, D. Reinkensmeyer and V. Dietz , eds., Springer, Cham, Switzerland, pp. 395–407.
Tan, A. Q. , and Dhaher, Y. Y. , 2014, “ Evaluation of Lower Limb Cross Planar Kinetic Connectivity Signatures Post-Stroke,” J. Biomech., 47(5), pp. 949–956. [CrossRef] [PubMed]

Figures

Grahic Jump Location
Fig. 1

Rendering of robotic gait trainer utilizing a single motor to produce a natural gait trajectory. The device can produce adaptable gait patterns by link adjustments.

Grahic Jump Location
Fig. 2

Mean of total 113 healthy subjects' gait pattern (meta-trajectory) and total nine gait patterns with different step lengths in x–y plane (average of ten subjects for large, medium, and small step lengths and their intermediates)

Grahic Jump Location
Fig. 3

Parameterized structure of the proposed mechanism. The end-effector path at PE is determined based on the angle of the rotary crank, θ2 and the configuration of 12 links with given lengths.

Grahic Jump Location
Fig. 4

Optimized to the meta-trajectory, the structure results in an RMS error of 3.09 cm: (a) crosses delineate the meta-trajectory, and circles show the predicted trajectory and (b) meta-trajectory and predicted endpoint trajectories in gait cycle domain with a constant crank (L1) input angular velocity

Grahic Jump Location
Fig. 5

Gait pattern error for number of adjustable links. Selected adjustable links are denoted on each bar graph. Adjusting just two links (L4 and L8) provides the largest drop in error for the fewest adjustments.

Grahic Jump Location
Fig. 6

The RMS error for each canonical gait pattern given two link adjustments shows consistent performance across patterns, with expectedly larger error in larger gait patterns

Grahic Jump Location
Fig. 7

(Left) Motorized wooden model of the gait trainer with accompanying range of motion reflecting a natural gait pattern. (Right) Experimental results using optical motion capture system by adjusting link lengths to match with three different step lengths (largest-top, medium-middle and smallest-bottom).

Grahic Jump Location
Fig. 8

Rendering of expected alpha prototype. The design of the gait trainer involves a robust frame with body weight support, a removable pelvic support, a treadmill, and the operational linkage mechanism, driven by a DC gear motor and belt drive.

Tables

Errata

Discussions

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging and repositioning the boxes below.

Related Journal Articles
Related eBook Content
Topic Collections

Sorry! You do not have access to this content. For assistance or to subscribe, please contact us:

  • TELEPHONE: 1-800-843-2763 (Toll-free in the USA)
  • EMAIL: asmedigitalcollection@asme.org